Measurement Systems International Inc DL7200RFL Benutzerhandbuch

Seite von 32
M E A S U R E M E N T  S Y S T E M S  I N T E R N A T I O N A L
Page 22     MSI-7200 Dyna-Link   
   User Guide
OPTIONS & ACCESSORIES
Top and bottom shackles (Crosby-Laughlin or equal)
Portable carrying case
Hard-wired remote control (single display, in remote)
Hard-wired remote control, dual display (displays in load cell and in
remote)
Rugged, outdoor rated, cable connectors, load cell to remote
External set-point outputs (2)
Remote output relay box for interfacing to set-points (requires external
set-point output option above)
External “New Peak” output
AC/DC power input (90-260 Vac / Vdc and/or 6 to 28 Vdc)
Analog output: nominally 0-2 volt equals zero load to full capacity
RS232 output (conforms to EIA232-D). Available with remote control
only
Dual load sensor input and summing
Quick disconnect for remote control options
SERVICE INITIALIZATION PROCEDURE
WARNING: This procedure is intended only for the use of MSI authorized
repair facilities only. The improper setting of capacity can cause a safety
hazard. Capacity is dictated by the MSI rated capacity of the Link load cell.
The meter electronics must not be set for a higher capacity than the Dyna-Link
is rated for.
Function
When a load cell or the electronics are in need of replacement, initialization of
the system is required. After the unit is “Cleared” a calibration must be per-
formed. Do not perform this procedure unless you are prepared to do a full cali-
bration with adequate test weights.
Uncalibrated Display
When the Version 3.12 Dyna-Link is not calibrated, the display will flash the
LEDs and blink the message “CAL.” This will stop when the Dyna-Link unit is
in setup, and when calibration has been successful.
Initialization Procedure
Standard Dyna-Links:
1) Remove the four seal screws holding on the back of the Dyna-Link.