Fukuda Denshi Co Ltd LX7230G Benutzerhandbuch

Seite von 101
6. SpO
Monitoring 
32 
4. 
If desired, the sensor cable can be secured to the patient’s clothing or 
other material by using the clip located on the cable. To open, pinch the 
sides of the clip; release to close (Figure (4)). 
 
(4) 
 
 
5. 
Connect the OXIMAX
®
 
MAX-FAST oximetry sensor into the LX-7230N. 
Verify proper operation as described in the operation manual. 
 
N O T E  
If the sensor does not track the pulse reliably, it may be incorrectly 
positioned - or the sensor site may be excessively wrinkled, or too deeply 
pigmented or otherwise deeply colored (for example, as a result of 
externally applied coloring such as dye or pigmented cream) to permit 
appropriate light transmission. If any of these situations occurs, reposition 
the sensor in a different location or choose an alternate Nellcor sensor to 
use on a different site. 
 
W A R N I N G  
  Precautions for Use of headband 
 
Do not use headband on children age 24 months and younger. 
 
Do not use headband on children with open fontanelles.
   
 
For details, refer to the instruction manual of OXIMAX
®
 MAX-FAST 
oximetry sensor.
 
 
C A U T I O N 
  Precautions for Use of headband 
 
Applying the headband too loose or too tight can cause inaccurate 
readings. Make sure the headband applies equal pressure to the entire 
sensor. The sensor must be completely covered by the headband. 
 
For details, refer to the instruction manual of OXIMAX
®
 MAX-FAST 
oximetry sensor.