Vernier Software & Technology LABQUEST2 Benutzerhandbuch
Software Updates – Software updates for LabQuest, as they are released, will be available at
www.vernier.com/labquest/updates. To determine what version of the software you are
using, tap System on the Home screen, then tap Information.
using, tap System on the Home screen, then tap Information.
Screen Calibration – The screen may need to be recalibrated periodically. To access the calibration
tool, tap System on the Home screen, then tap Calibrate Screen. If you are unable to access
the Calibration tool, press and hold the power button until the LabQuest shutdown message is
displayed. Wait until the system shuts down then turn the unit back on. While the system is
rebooting, press the Home key to prompt the system to display the calibration tool. Tap the
calibration target in the five locations indicated to calibrate the screen.
the Calibration tool, press and hold the power button until the LabQuest shutdown message is
displayed. Wait until the system shuts down then turn the unit back on. While the system is
rebooting, press the Home key to prompt the system to display the calibration tool. Tap the
calibration target in the five locations indicated to calibrate the screen.
System Reset – To reset the system, tap System on the Home screen, then tap Reboot. If you are
unable to access the Reboot tool, press and hold the power button until the LabQuest
shutdown message is displayed. Wait until the system shuts down then turn the unit back on.
shutdown message is displayed. Wait until the system shuts down then turn the unit back on.
Battery Life – Battery life will depend on the sensors and features used. To access power saving
options, tap Preferences on the Home Screen, then tap Light and Power.
Battery Maintenance – Use only the supplied AC adapter or optional charging station to charge the
LabQuest battery. A full charge can take eight hours. The battery cannot be overcharged,
and it can be safely recharged after a partial discharge.
and it can be safely recharged after a partial discharge.
Screen Maintenance – The LabQuest screen is water resistant. Wipe the screen clean with a cotton
cloth that is slightly dampened with water or ethanol. Do not use any other solvents. Do not
submerge your LabQuest in liquids.
submerge your LabQuest in liquids.
Summer Storage – For long-term storage of your LabQuest, fully shutdown the unit by pressing and
holding the power button until the shutdown message is displayed. After the system shuts
down you can remove the battery, if desired. Store the LabQuest in a climate-controlled
environment. Excessive heat can significantly reduce battery life.
down you can remove the battery, if desired. Store the LabQuest in a climate-controlled
environment. Excessive heat can significantly reduce battery life.
Periodic Table
20
Other Applications
21
Other Applications
Care of LabQuest
Care of LabQuest
22
23
Audio Function
Generator
Generator
Periodic Table, Audio Function
Generator, and Power Amplifier
(not shown) are accessed via the
Home menu.
Generator, and Power Amplifier
(not shown) are accessed via the
Home menu.
Stopwatch, Calculator, and Sound
Recorder (not shown) are accessed
via the Home> Accessories menu.
Recorder (not shown) are accessed
via the Home> Accessories menu.
Calculator
Stopwatch