Operator’s Manual  ZM-520PA/521PA/530PA/531PA
Electromagnetic Immunity
This Model ZM-520PA/530PA is intended for use in the electromagnetic environment specified 
below. The customer or the user of the ZM-520PA/530PA should assure that it is used in such an 
environment.
Immunity test
IEC 60601 test 
level
Compliance 
level
Electromagnetic 
environment - guidance
Electrostatic 
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete 
or ceramic tile. If floors are 
covered with synthetic material, 
the relative humidity should be 
at least 30%.
Electrical fast 
transient/burst
IEC 61000-4-4
±2 kV for power 
supply lines
±1 kV for input/
output lines
Not applicable
Surge
IEC 61000-4-5
±1 kV differential 
mode
±2 kV common mode
Not applicable
Voltage dips, short 
interruptions and 
voltage variations 
on power supply 
input lines
IEC 61000-4-11
<5% 
U
T
 (>95% dip in 
U
T
) for 0.5 cycles
40% 
U
T
 (60% dip in 
U
T
) for 5 cycles
70% 
U
T
 (30% dip in 
U
T
) for 25 cycles
<5% 
U
T
 (>95% dip in 
U
T
) for 5 s
Not applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields 
should be at levels characteristic 
of a typical location in a 
typical commercial or hospital 
environment.
NOTE:  
U
T
 is the AC mains voltage prior to application of the test level.
Avoiding Electromagnetic Interference (Impedance Respiration)
Impedance respiration measurement is very sensitive and affected by electromagnetic 
interference. Technological limitations do not allow immunity levels higher than 1 V/m for 
radiated RF electromagnetic fields. Electromagnetic fields with field strengths above 1 V/m 
may cause measurement error. Do not use electrically radiating equipment near the impedance 
respiration measurements.