Operator’s Manual  ZM-520PA/521PA/530PA/531PA
 17
CAUTION
If the attachment site is dirty with blood 
or bodily fluids, clean the attachment 
site before attaching the probe. If there 
is nail polish on the attachment site, 
remove the polish. Otherwise, the 
amount of transmitted light decreases, 
and measured value may be incorrect 
or measurement cannot be performed.
CAUTION
If the skin gets irritated or redness 
appears on the skin from the probe, 
change the attachment site or stop 
using the probe. Take extreme care for 
the patients with delicate skin.
CAUTION
Failure to follow these instructions may 
cause cable discontinuity, short circuit, 
skin burn on the patient from the probe 
temperature increase due to the short 
circuit of the probe cable, and incorrect 
measurement data. If the probe is 
broken, replace it with a new one.
•  Do not immerse any part of 
the probe cable other than the 
disposable probe in chemical 
solutions or water. 
•  Do not pull or bend the probe cable.
•  Do not let caster feet run over the 
probe cable.
CAUTION
When the probe is attached on an 
appropriate site with sufficient 
circulation and the error message 
confirming the probe attachment 
repeatedly appears, the probe may be 
deteriorated. Replace it with a new 
one.
Maintenance
CAUTION
This transmitter is not waterproof. If 
detergent or liquid spills into the 
transmitter, stop cleaning or 
disinfecting it and contact your Nihon 
Kohden representative. The transmitter 
needs to be checked for safety and 
function before use.
CAUTION
Do not immerse the electrode lead 
connector in liquid.