Operator’s Manual  ZM-520PA/521PA/530PA/531PA
 79
Channel numbers
Startup screen
Numeric and waveform screen
3. Transmitting/Receiving  Signal
Use the AX-400G vital sign simulator and receiving monitor.
1.  Connect the vital sign simulator to the transmitter.
AX-400G
2.  Place the transmitter 2 to 3 m from the receiving monitor.
3.  Set the channel on the receiving monitor to the channel of the transmitter.
4.  Turn on the transmitter and vital sign simulator.
5.  Check that the ECG of the transmitter appears on the receiving monitor.
6.  Turn off the transmitter.
7.  Check that the ECG disappears from the receiving monitor.
4. Display
Check that there are no dots missing on the screen.
1.  Turn off the transmitter.