Operator’s Manual  ZM-520PA/521PA/530PA/531PA
5.  Press each key one at a time and check that the pressed key is highlighted on the screen.
When the CALL key is 
pressed, the key name is 
highlighted.
6.  After checking, press and hold the Screen key to return to the MANUAL CHECK screen.
6. ECG Check
1.  Connect the vital sign simulator to the transmitter.
AX-400G
2.  Place the transmitter 1 m from the receiving monitor.
3.  Turn on the transmitter and vital sign simulator.
4.  Check that the ECG of the transmitter appears on the receiving monitor.