Operator’s Manual  ZM-520PA/521PA/530PA/531PA
Output Signal
WARNING
Do not use the output signal from the receiving monitor as the synchronization signal 
for other equipment such as IABP, MRI, echocardiography or defibrillator. There may 
be time delay between the monitor and the other equipment caused by waveform 
transmission delay and spike noise may interfere on the output signal and be 
mistaken as a trigger.
Battery
WARNING
Do not use NiMH batteries for this 
transmitter. Some operating 
environments may cause NiMH 
batteries to produce gas and explode.
CAUTION
Battery replacement must be 
performed by the operator. When 
replacing the batteries of a transmitter 
that is currently used for a patient, 
disconnect the electrode leads from 
the transmitter before replacing 
batteries or do not touch the patient 
during replacement.
CAUTION
When the transmitter is not in use, 
remove the batteries. When the 
batteries are installed, battery power is 
consumed even when measurement is 
not performed.
CAUTION
Do not handle the batteries with wet 
hands.
CAUTION
Refer to the battery manual for details 
on handling the batteries.