Operator’s Manual  ZM-520PA/521PA/530PA/531PA
WARNING
When not monitoring SpO
2
, disconnect 
the SpO
2
 cable from the transmitter. 
Otherwise, noise from the probe 
sensor may interfere and incorrect data 
is displayed on the screen.
CAUTION
Refer to the probe instruction manual 
for details.
CAUTION
The disposable probe is not sterilized. 
Use the disposable probe only for a 
single patient. Never reuse the 
disposable probe for another patient 
because it causes cross infection.
CAUTION
Do not attach the probe to the same 
limb that is used for NIBP 
measurement or an IBP catheter. The 
SpO
2
 measurement may be incorrect.
CAUTION
While a patient is on medication which 
causes vasodilation, the pulse 
waveform may change and in rare 
cases the SpO
2
 value might not be 
displayed.
CAUTION
Normal external light does not affect 
monitoring but strong light such as a 
surgical light or sunlight may affect 
monitoring. If affected, cover the 
measuring site with a blanket.
CAUTION
When a message indicates a faulty 
probe, stop monitoring and replace the 
probe with a new one.
CAUTION
Do not use a damaged or 
disassembled probe. It causes 
incorrect measurement and may injure 
the patient.
CAUTION
Do not use a probe which is 
deteriorated by aging. Accurate 
measurement cannot be performed.