Operator’s Manual  ZM-520PA/521PA/530PA/531PA
 49
CAUTION
If the attachment site is dirty with blood 
or bodily fluids, clean the attachment 
site before attaching the probe. If there 
is nail polish on the attachment site, 
remove the polish. Otherwise, the 
amount of transmitted light decreases, 
and measured value may be incorrect 
or measurement cannot be performed.
CAUTION
When the probe is attached on an 
appropriate site with sufficient 
circulation and the error message 
confirming the probe attachment 
repeatedly appears, the probe may be 
deteriorated. Replace it with a new 
one.
CAUTION
If the skin gets irritated or redness 
appears on the skin from the probe, 
change the attachment site or stop 
using the probe. Take extreme care for 
the patients with delicate skin.
CAUTION
Do not use a probe which is 
deteriorated by aging. Accurate 
measurement cannot be performed.
CAUTION
Refer to the probe instruction manual 
for details.
CAUTION
The disposable probe is not sterilized. 
Use the disposable probe only for a 
single patient. Never reuse the 
disposable probe for another patient 
because it causes cross infection.