Operator’s Manual  ZM-520PA/521PA/530PA/531PA
Turning ECG Measurement On/Off
ECG measurement can be turned on or off on the PARAMETER SETUP screen. When electrodes 
are attached to the patient and ECG leads are connected, ECG monitoring starts even when ECG 
is turned off.
When PROTOCOL on the SYSTEM SETUP screen is set to 51:
ECG measurement on the receiving monitor is automatically set to off.
NOTE
ECG measurement on the transmitter cannot be turned on or off from the receiving 
monitor.
When PROTOCOL on the SYSTEM SETUP screen is set to 41:
If ECG measurement is turned off on the transmitter, ECG measurement on the receiving monitor 
must also be turned off.
Turning Respiration Measurement On/Off
Respiration measurement can be turned on or off on the PARAMETER SETUP screen. If 
respiration measurement is turned off, respiration measurement on the receiving monitor is also 
turned off.
Electrode Detachment
In the following conditions, the “CHECK ELECTRODES” message is displayed on the 
transmitter and receiving monitor.
•  Electrode is detached from skin.
•  Electrode lead is disconnected from the electrode.
•  Polarization voltage between the electrode and skin is excessively high.
In these cases, check the cause and if necessary, replace electrodes with new ones.
CAUTION
When the “ELECTRODE OFF” or “CHECK ELECTRODES” message is displayed on 
the receiving monitor, ECG is not monitored properly and the ECG alarm does not 
function. Check the electrode, electrode leads, and if necessary, replace with new 
ones.