Operator’s Manual  ZM-520PA/521PA/530PA/531PA
Maintenance
To use the transmitter in safe and optimum condition, perform maintenance check every six 
months.
The following units are necessary for some checking items.
•  AX-400G Vital Sign Simulator
• AX-300T SpO
2
 Checker
• Receiving monitor
CAUTION
Never disassemble or repair the transmitter. If there is any problem with the 
transmitter, contact your Nihon Kohden representative.
NOTE
•  The measurement accuracy of the above units must be managed to perform accurate 
maintenance check.
•  For details on the operation of the above units, refer to the manuals provided with 
these units.
A maintenance check sheet is provided at the end of this section. Make a copy of this check sheet 
before performing maintenance check.
1. External Check
•  There is no damage or dirt on the outside of the transmitter.
•  The battery case cover is not damaged, the spring is firmly attached and the battery case cover 
can be closed firmly.
•  No keys are damaged.
•  No electrode leads are damaged.
•  There are no blood or chemicals on the transmitter.
•  The springs in the battery compartment are not damaged or detached.
•  Terminals in the battery case are not corroded.
2. Transmitter  Channel
•  The channel number label on the transmitter is not torn or removed.
•  The channel of the transmitter matches the label.
  The transmitter channel is displayed in the upper left corner of the screen. The channel number 
also appears on the startup screen.