Operator’s Manual  ZM-520PA/521PA/530PA/531PA
 83
7. Respiration Check
1.  Connect the vital sign simulator to the transmitter.
AX-400G
2.  Place the transmitter 1 m from the receiving monitor.
3.  Turn on the transmitter and vital sign simulator.
4.  Check that the respiration waveform of the transmitter appears on the receiving monitor.
8. SpO
2
 Check
1.  Connect the SpO
2
 checker to the transmitter.
AX-300T
2.  Place the transmitter 1 m from the receiving monitor.
3.  Turn on the transmitter and SpO
checker.
4.  Check that the pulse bar graph appears on the transmitter screen.