Operator's Manual  ZM-920PA/930PA
CAUTION
• Battery replacement must be performed by the operator.  When
replacing batteries of the transmitter currently used for a patient,
disconnect electrode leads from the transmitter before replacing
batteries or do not touch the patient during replacement.
• The battery charger must be used outside the patient environment.
• Refer to the battery and battery charger manuals for details on
handling the batteries.
WARNING
• Keep the battery pack away from fire.  It may explode.
• Do not damage, disassemble, drop or give impact to the battery.  If
the battery is damaged and substance inside the battery contacts
the skin or clothes, wash immediately and thoroughly with water.
• Never short-circuit the + and – terminals on the battery.  It may
cause overheating and fire.
• Take care that the patient does not swallow batteries.
Battery
Transmitter Channel Management
WARNING
The following actions must be taken to properly receive the
transmitter signal of the correct patient on the receiving instrument,
otherwise, there may be signal loss or signals may mix causing a
serious accident, such as monitoring a different patient.
••••• Assign a channel administrator in the hospital and only he or she
should manage channel assignments.
••••• The channel administrator must manage the channels in the
facility so that there is no signal interference.
••••• When the transmitter channel is changed, the channel
administrator must check that the channel on the receiving
monitor is also changed and the signal is properly received.