ECG Monitoring
CAUTION
• Only use Nihon Kohden specified electrodes and electrode leads.
When other type of electrodes or electrode leads are used, the
“CHECK ELECTRODE” message may be displayed and monitoring
may stop.
• When the “ELECTRODE OFF” or “CHECK ELECTRODE” message
is displayed on the receiving monitor, ECG is not monitored
properly and the ECG alarm does not function.  Check the
electrode and electrode leads, and if necessary, replace with new
ones.
SpO
2
 Monitoring
WARNING
• SpO
2
 measurement may be incorrect in the following cases.
• When the patient’s carboxyhemoglobin or methemoglobin
increases abnormally.
• When dye is injected in the blood.
• When using an electrosurgical unit.
• During CPR.
• When measuring at a site with venous pulse.
• When there is body movement.
•  When the pulse wave is small (insufficient peripheral
circulation).
• Check the circulation condition by observing the skin color of the
measuring site and pulse waveform.  Change the measuring site
every 8 hours for disposable probes and every 4 hours for
reusable probes.  The skin temperature may increase at the
attached site by 2 or 3°C (4 or 5°F) and cause a burn or pressure
necrosis.  When using the probe on the following patients, take
extreme care and change the measurement site more frequently
according to symptoms and degree.
• A patient with a fever