Nihon Kohden Corporation ZM-940PA Benutzerhandbuch

Seite von 111
Operator's Manual  ZM-940PA
1
Introduction
The ZM-940PA transmitter transmits ECG, respiration, SpO
2
, NIBP and pulse waveform from
a patient to a Nihon Kohden monitor for continuous monitoring.  The transmitter can change
channels when connected to the QI-901PK channel writer.  The front LCD displays SpO
2
%,
NIBP, pulse rate, pulse waveform amplitude, electrode condition mark, battery condition and
NIBP measuring mode and interval.
Read the operator’s manual for the receiving monitor together with this manual before operation.
CAUTION
• Do not use the same channel for different patients.  If the same channel is
used for two patients, the two patients’ data will be lost due to mutual
modulation interference, or another patient’s data may appear on the receiving
monitor screen.
• Do not use transmitters of adjacent channels in a hospital.  If a transmitter of
an adjacent channel is used, radio waves from one transmitter affect the
receiver of the adjacent channel’s transmitter and there may be interference.
• Do not use the same transmitter on more than one patient at the same time.
Do not connect different sensors on different patients to the same transmitter.
NOTE
• To prevent interference between channels, assign a channel administrator in
the hospital and only he or she should manage channel assignment.
• Use Nihon Kohden parts and accessories to assure maximum performance
from your instrument.
• For stable signal reception, it is recommended to use a diversity antenna
system on the receiving monitor.  Otherwise, spike noise from transient fading
of electric field strength (for example, people moving) may interfere with the
transmitter signal and may be mistaken as an arrhythmia on the receiving
monitor.
• NIBP cannot be measured on a neonate. (ECG, respiration and SpO
2
 can be
monitored on a neonate.)
• For details on the receiving monitor and upgrade information, contact your
Nihon Kohden representative.