Operator's Manual  ZM-940PA
Notes on Parameter Settings
When monitoring NIBP and SpO
2
, the following setting must be set as indicated in the table to
properly transmit the monitoring data to the receiving monitor.  Otherwise, SpO
2
 cannot be
monitored properly during NIBP measurement.
Some receiving monitors require the software to be upgraded.  For details, contact your Nihon
Kohden representative.
SpO
2
 probe attachment site
INHIBIT SpO
2
 DURING NIBP setting
Probe attached to the same limb as the cuff
ON
Probe attached to the limb without cuff*
OFF
* When the SpO
2
 probe is attached to the same limb as the NIBP cuff and the cuff is
inflated, the SpO
2
 value becomes unstable and SpO
2
 or PR alarm may occur.