Operator's Manual  ZM-940PA
tape causes incorrect measurement and may injure the skin at the attachment
site.
• When fixing the TL-260T multi-site Y probe with the sponge attachment tape,
confirm that the adhesive part of the tape is not on the skin.  The adhesive may
cause oversensitive symptoms on the skin such as redness or itch.  If the
adhesive touches the skin, remove it carefully and slowly because neonatal
skin is very delicate.
• Do not use a dirty sponge attachment tape.  The measurement value may be
incorrect.
• When removing the probe from the attachment tape, do not pull the sensor
cable because this can damage the cable.
• Use the sponge attachment tape S and L only with the specified products.
• Refer to the probe instruction manual for details.
* Monitoring SpO
2
 during NIBP Measurement
When the SpO
2
 probe is attached to the same limb as the NIBP cuff, the blood flow decreases
during NIBP measurement and pulse wave cannot be detected and SpO
2
 cannot be monitored
properly.   When “INHIBIT SpO
2
 DURING NIBP” on the PARAMETER SETUP screen is
set to ON (factory default setting), SpO
2
 monitoring is paused during NIBP measurement to
avoid SpO
2
 alarm occurrence.  However, when monitoring SpO
2
 on the same limb as the
NIBP, be careful when reading SpO
2
 values.
Maintenance
WARNING
• Before cleaning or disinfecting the transmitter, remove the batteries from the
transmitter.
• The transmitter cannot be sterilized.
CAUTION
• This transmitter is not waterproof.  If detergent or liquid spills into the
transmitter, stop cleaning/disinfecting/sterilizing it and contact your Nihon
Kohden representative.  The transmitter needs to be checked for safety and
function before use.
• Never disassemble or repair the transmitter.  When there is any problem with
the transmitter after maintenance and inspection, contact your Nihon Kohden
representative.