Operator's Manual  ZM-940PA
29
Changing System Setup Settings
NOTE
Changing System Setup settings must be done only by a qualified personnel.
System Setup Setting List
The factory default settings are underlined.
Setting Item
Description
Settings
CHANNEL
Displays the transmitter channel.
PRESSURE UNIT
Select the units for NIBP.
mmHg, kPa
LANGUAGE
Select the language for screen display.
JPN, ENG
BRIGHTNESS
Select the LCD brightness.
1, 2, 3, 4
SYSTEM
INITIALIZE
Initializes all settings to the factory default
settings.
Displaying the SYSTEM SETUP Screen
1.
Remove one battery.
2.
While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery.
The SETUP screen appears.
3.
Press the NIBP INTERVAL key to move the cursor to “SYSTEM SETUP”.
4.
Press the NIBP START/STOP key to enter the
SYSTEM SETUP screen.
When the cursor is moved to “EXIT” by
pressing the NIBP INTERVAL key and the
NIBP START/STOP key is pressed, the startup
screen appears, then the monitoring screen
appears.