Operator's Manual  ZM-940PA
or pressure necrosis from poor blood circulation, especially on neonates or
low birth weight infants whose skin is delicate.  Accurate measurement cannot
be performed on a site with poor peripheral circulation.
• Check the circulation condition by observing the skin color of the measuring
site and the pulse waveform.  Change the measuring site every 8 hours for
disposable probes and every 4 hours for reusable probes.  The skin
temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and
cause a burn or pressure necrosis.  When using the probe on the following
patients, take extreme care and change the measurement site more
frequently according to symptoms and degree.
• A patient with a fever
• A patient with a peripheral circulation insufficiency
• Neonate or low birth weight infant with delicate skin
CAUTION
• If the attachment site is dirty with blood or bodily fluids, clean the attachment
site before attaching the probe.  If there is nail polish on the attachment site,
remove the polish.  Otherwise, the amount of transmitted light decreases and
measured value may be incorrect or measurement cannot be performed.
• If the skin gets irritated or redness appears on the skin from the probe,
change the attachment site or stop using the probe. Take extreme care for the
patients with delicate skin.
• When the probe is attached on an appropriate site with sufficient circulation,
but the error message confirming the probe attachment repeatedly appears,
the probe may be deteriorated.  Replace it with a new one.
• Do not use a probe that is deteriorated by aging.  Accurate measurement
cannot be performed.
• Neonatal skin is delicate. Remove the probe and tape carefully and slowly.
• When removing a probe that is taped to the skin, do not pull the probe cable
because this can damage the cable.
• When removing the probe from the attachment tape, do not pull the sensor
cable because this can damage the cable.
• Do not immerse the disposable probe in detergents or water. If the probe
adhesive surface gets wet, adhesiveness becomes weak and the probe
cannot be attached to the skin.
• Before using the TL-260T multi-site Y probe, be sure to attach the probe to the
sponge attachment tape S or L.  Do not use the probe without the sponge
attachment tape attached.  Using the probe without the sponge attachment
tape causes incorrect measurement and may injure the skin at the attachment
site.