Nihon Kohden Corporation ZM-940PA Benutzerhandbuch

Seite von 111
Operator's Manual  ZM-940PA
63
Electrode Detachment
In the following conditions, the check electrode indication is displayed on the LCD of the
transmitter and the “CHECK ELECTRODE” message is displayed on the monitor.
• Electrode is detached from skin.
• Electrode lead is disconnected from the electrode.
• Polarization voltage between the electrode and skin is excessively high.
In these cases, check the cause and if necessary, replace electrodes with new ones.
CAUTION
When the “ELECTRODE OFF” or “CHECK ELECTRODE” message is displayed
on the receiving monitor, ECG is not monitored properly and the ECG alarm dose
not function. Check the electrode and electrode leads and if necessary, replace
with new ones.
Check Electrode Indication on the Transmitter when Monitoring with 3 Electrodes
The “
” mark is displayed.
Check electrode symbol
Check Electrode Indication on the Transmitter when Monitoring with 6 Electrodes
The “
” mark and either the detached lead or detached electrode position is indicated,
depending on the LEADS OFF DISPLAY setting on the PARAMETER SETUP screen.
Check electrode indication
When LEADS OFF
DISPLAY is set to CHAR, the
detached lead is indicated
When LEADS OFF
DISPLAY is set to IMAGE,
the detached electrode
position is indicated with X
Check electrode symbol