Teguar Corporation TMT-4391-08 Benutzerhandbuch

Seite von 12
 
Recommended Separation Distances Between
 
Portable and Mobile RF Communications Equipment 
and the TMT-4391-08
 
 
TMT-4391-08 is intended for use in an electromagnetic 
environment in which radiated RF disturbances are 
controlled. The customer or the user of the model 
TMT-4391-08 can help prevent electromagnetic 
interference by maintaining a minimum distance between 
portable and mobile RF communications equipment 
(transmitters) and the model TMT-4391-08 as 
recommended below, according to the maximum output 
power of the communications equipment. 
 
Rated 
Maximum
 
 
Output 
Power of
 
Separation Distance According to 
Frequency of Transmitter
 
Transmitter  m 
 
 
W 
 
 
 
 
 
 
 
 
150 kHz to 
80 MHz
 
80 MHz to 
800 MHz
 
800 MHz to 
2,5 GHz
 
 
d = 1,2  P  d = 1,2  P  d = 2,3  P 
 
 
 
 
0,01 
0,12 
0,12 
0,23 
 
 
 
 
0,1 
0,38 
0,38 
0,73 
 
 
 
 
1,2 
1,2 
2,3 
 
 
 
 
10 
3,8 
3,8 
7,3 
 
 
 
 
100 
12 
12 
23 
 
 
 
 
For transmitters rated at a maximum output power not 
listed above, the recommended separation distance d in 
meters (m) can be estimated using the equation in the 
table above applicable to the frequency of the transmitter, 
where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer. 
 
NOTE 1 At 80 MHz and 800 MHz, the separation 
distance for the higher frequency range applies. 
 
NOTE 2 These guidelines may not apply in all situations. 
Electromagnetic propagation is affected by absorption 
and reflection from structures, objects and people. 
 
 
Guidance and Manufacturer’s Declaration – 
Electromagnetic Immunity
 
 
TMT-4391-08 is intended for use in the electromagnetic 
environment specified below. The customer or the user 
of the model TMT-4391-08 should assure that it is used 
in such an environment. 
 
 
 
Immunity 
Test
 
IEC 60601 
Test
 
Compliance 
Electromagnet
ic 
Environmental
  
Level 
Level 
Guidance 
 
 
 
 
 
 
 
 
Electrostatic 
dis- 
 
 
Floors should 
be wood, con-   
±
8 kV 
contact 
±
8 kV 
contact 
crete or ceramic 
tile. If floors 
 
charge (ESD) 
are covered 
with synthetic 
 
±
15 kV 
air 
±
15 kV air 
 
IEC 
61000-4-2 
material, the 
relative 
humidity 
 
 
 
 
 
 
 
should be at 
least 30%. 
 
 
 
 
 
 
Electrical fast 
±
2 kV for 
power 
±
2 kV for 
power 
 
 
supply lines  supply lines 
Main power 
quality should 
be 
 
transient/bur
st 
 
 
 
that of a typical 
commercial or 
 
 
±
1 kV for 
input/out- 
±
1 kV for 
input/ 
 
IEC 
61000-4-4 
hospital 
environment. 
 
put lines 
output lines 
 
 
 
 
 
 
 
 
 
 
 
±
1 kV 
line(s) to 
±
1 kV 
line(s) to 
Main power 
quality should 
be 
 
Surge 
line(s) 
line(s) 
 
that of a typical 
commercial or 
 
IEC 
61000-4-5 
±
2 kV 
line(s) to 
±
2 kV 
line(s) to 
 
hospital 
environment. 
 
 
earth 
earth 
 
 
 
 
 
 
 
 
 
 
<5% U
(>95% dip 
<5% U
 
 
 
(>95% dip in   
 
 
in UT) for 
0,5 cycle 
UT) for 0,5 
Main power 
quality should 
be 
 
 
 
cycle 
 
 
40% U
 
that of a typical 
commercial or   
Interruptions 
and 
(60% dip in 
UT) 
40% U
(60% 
hospital 
environment. If 
the 
 
voltage varia- for 5 cycles 
dip in UT) for 
user of the 
model 
TMT-4391-08   
tions on 
power 
 
cycles 
requires 
continued 
operation 
 
supply input 
lines 
70% U
 
during main 
power interrup-   
 
(30% dip in 
UT) 
70% U
(30% 
tion, it is 
recommended 
that 
 
IEC 
61000-4-11 
for 25 
cycles 
dip in UT) for 
the model 
TMT-4391-08 
be 
 
 
 
25 cycles 
powered from 
an uninterrupt-