Electrical hazard and electric shock risk 
 
CAIs and CAIx connector should not be in contact with any other conductive components 
including conductive material and ground connection. 
 
CAI has a potential to develop an error in vital signals to be obtained when the device is 
located around heart electric defibrillator or a device giving electric shock. 
 
To prevent vital signal error of this device and ensure operation stability, do not use the device 
together with heart electric defibrillator or a device to giving high electric shock. 
 
Shock risk: Do not attempt to connect or disconnect CAIs and CAIx with wet hand. When 
connecting CAIs, check that the user’s hands are clean and dry prior to use.  
 
Risk of electric shock: Do not remove CAIx cover when the CAI monitor is being operated or 
plugged in to the power outlet. 
 
Electrical Shock Risk and Safety Prevention:  Although this device was manufactured with 
ground leakage current and patient safety current adequate to applicable standard safety 
specification, the device should receive regular inspection at least once a year for electrical 
safety prevention. 
 
When blood or any other liquid is spilled on this device, do not continue to use it and must 
conduct ground leakage current test immediately. 
 
Power supply is equipped with an internal fuse and when replacement of power supply is 
required, it should be replaced with new power supply comprising CAI device of the 
manufacturer. 
 
This device is a power supply embedded device, which has rechargeable battery in the device. 
The battery has overcharge prevention and its replacement should be made through the 
manufacturer.  
Caution: Do not disassemble or replace manually except by qualified personnel from the 
manufacturer.   
 
This device is a power supply embedded product, which includes no equipotential access. 
 
Caution: This device is a power supply embedded product, so do not connect and use several 
instruments simultaneously due to risks that might be brought by summing of leakage current.