CORITICAL ACTIVITY INDEX 
|
 17
 
 
 
• 
Bluetooth function should be turned on only when using CAI. 
• 
When any suspicious malignant code or virus was invaded into CAI monitor or the 
initialization of device is required, inquire to customer service or local representative of 
Charm Engineering. 
• 
When storage space of device is not enough, stored data should be backed up to 
external or other storage space for future use. 
 
Cleaning 
When cleaning and sterilizing CAI, do not autoclave it. Autoclaving can damage most of CAI 
components. 
 
Prevention of Electromagnetic Interference  
 
Although CAI satisfies electromagnetic compatibility requirements of IEC 60601-1-2, operation 
of CAI may give or receive effects due to electromagnetic interference (EMI) of other device. 
 
To prevent electromagnetic interference (EMI), 
• 
Increase distance between devices.  
• 
Change location of power line and cable of the device. 
• 
CAI monitor should maintain at least 15 cm of distance from other device to minimize 
EMI. 
• 
As CAI device may be affected by operation of portable mobile radio frequency 
communication device, it is not recommended to use mobile communication device.  
 
CAI should not be placed beside or on top of other device during usage. 
When it is unavoidable to use it near to other devices, ensure to check whether CAI operates 
as in normal conditions prior to usage. 
 
※ Use of any accessary or component not mentioned specifically in this user manual, can 
increase electromagnetic discharge of CAI and decrease electromagnetic immunity. 
 
※ Maintenance, repair, component replacement, and S/W update should be completed by the 
qualified medical engineers. 
 
※ CAI product of Charm Engineering was designed and manufactured to comply with medical 
device standard test of the Food and Drug Administration and provided as suitable device.