compartment of the XRpad™ 4336 MED detector using a slow and steady pull motion, 
supporting both the XRpad™ 4336 MED detector, and the XRpad™ LBP.  
 
Store the XRpad™ LBP in a cool, dry, clean environment if not in use or during recharge of 
the XRpad™ LBP for the next use. 
 
 
 
Figure 5  Removal of the XRpad™ LBP 
8.5.1.4
 
Transportation and Storage 
 
Store the XRpad™ LBP in a cool, dry, clean environment when not in use.  Do not remove 
the XRpad™ LBP from its original packaging until it is required for use. 
 
Do not leave, expose or store the XRpad™ LBP in extremely hot or cold temperatures (e.g., 
in direct sunlight, nearby heat sources, in cars or car trunks).   The XRpad™ LBP may 
overheat causing fire, or performance life will be shortened. 
 
Do not short-circuit the XRpad™ LBP, or store the XRpad™ LBP without sufficient 
packaging in a location where it may be short-circuited. 
8.5.1.5
 
Maintenance of the XRpad™ LBP 
 
Before inserting XRpad™ LBP into the XRpad™ 4336 MED detector or XRpad™ LBC 
battery charger, inspect the XRpad™ LBP for sign of damage, defects or abnormality.  Do 
not use damaged, defective or abnormal condition XRpad™ LBP.   
 
Check to ensure the XRpad™ LBP is clean, dry and free of foreign contamination or debris. 
If cleaning is necessary, refer to section 7.5.1.6 for cleaning instructions. 
 
The XRpad™ LBP has no repairable parts. Do not disassemble.  No modification of this 
product is allowed. 
 
If the XRpad™ LBP gives off an odor or generates heat or in any way appears abnormal 
during use, recharging or storage, immediately remove it from the device or battery charger 
and stop using the XRpad™ LBP.   
 
Using a damaged or defective XRpad™ LBP may cut operating time or cause the XRpad™ 
4336  MED detector system to fail. 
 
If a XRpad™ LBP leaks, do not touch the leaking fluid.  If the fluid touches your skin or 
eyes, wash the affected area with clean running water and immediately seek medical 
attention.  
 
If the XRpad™ LBP has not been used or charged for an extended amount of time 
(approximately 30 days), check the condition of the XRpad™ LBP and recharge if necessary 
prior to use. 
8.5.1.6
 
Cleaning  of the XRpad™ LBP 
 
Avoid exposure of the XRpad™ LBP to liquids and solvents when possible.  
 
Do not allow liquids or solvents to contact the electrical contacts on the XRpad™ LBP. 
 
When necessary, the XRpad™ LBP may be clean using a lightly moistened cloth with 70% 
isopropyl alcohol or 3% hydrogen peroxide. 
 
Never use thinner, benzene, acetone or any other corrosive or flammable cleaning agents. 
 
Ensure the XRpad™ LBP is completely clean and dry prior to storage, inserting into the 
XRpad™ 4336 MED detector or XRpad™ LBC battery charger.