Caution:
Do not take measurement during charging
Quick Start Guide
Upper Arm Blood Pressure Monitor
Please read this instruction before use this monitor
2
4
3
Thanks for buying our Upper Arm Blood Pressure Monitor 
【The important Instructions should be learned before use】
Systolic pressure
Under the table of Warning and Caution glyph is mainly to 
alert the user to the proper and safe use of this product, to 
prevent damage caused by misuse or harm the user and 
others.  
graphic 
E L B O W
SHOULDER
Ver1.0
【Structure and Instruction】
1
【Upper Arm Blood Pressure Monitor】
Cuff
Main unit
Take off the USB rubber cover, then you'll find 
Power/Function Button
MX-724BT
【The important Instructions should be learned before use】
【The important Instructions should be learned before use】
Note: please make the USB rubber cover well kept.
the USBinterface for charging
【Display】
indicator display
Diastolic pressure 
indicator display
Pulse indicator
display
Press unit
Pulse 
symbol
Pulse unit
Battary status 
indicator
Instruction of Symbol
content
Caution Indicates wrong use may cause casualties
Notice
Indicates wrong use may cause casualties
Indicates a general prohibition (which does not 
Indicates a general mandatory (that must comply 
Symbol Means BF Application
Symbol Means class II equipment
Electronic Information Products Pollution control 
protection use period of the product is 10 years.
Product Application
Suitable for: at least 12 years of age or older people 
Usage: Measuring blood pressure (systolic, diastolic) 
Positioning the Pressure Cuff
Positioning the pressure cuff: “          ”Positioning the 
The blood pressure monitor can be measured in either 
Approximately 1-2cm from the inside of elbow to the 
Fastness of the cuff and arm: The max gap between the 
arm and cuff are not more than one finger width; there
 is no gap between the cuff and arm when most tight, but 
comfortable cuff is wrapped around the arm.
edge of the cuff
upper arm, however, since there is a little difference in 
blood pressure between the left and right arms, so be 
sure to always use the same arm for measurement.
pressure cuff, “          ”Symbol toward the shoulder. 
and pulse
(does not apply to newborns)
symbol
allow things)
with the thing)
Symbol, indicates safety and environmental 
Place the blood pressure monitor display in front of your 
eyes, then you can see it easily.
Proper posture for taking a measurement
Stretch your back
Sit on the chair
Take off your coat, 
sweater or thick clothes, 
bare your upper arm or 
leave the thin clothes.
Make the centre of the 
 pressure cuff at the same 
height as your heart 
The best 
height of 
chair 
between 
table is 
25-30cm
Note
Take measurements while sit in a relax and quiet 
When measuring elbows on the table, palms up.
NoteIn: the following situations, accurate measurements 
Immediately after exercising or taking a bath; In cold 
Within 1 hour of smoking or drinking alcohol, coffee 
Incorrect posture, unstable mood during the 
Taking, moving body, arm, or bend fingures during 
Using mobile phone or other sources of 
Taking measurement continuously (between two 
Arrhythmia patients may have a greater impact on 
NoteIn: For the measurement results do not overly rely 
on their own judgment, be sure to accept the 
doctor's guidance.
environment;  before taking measurements, take 2 or 3 
deep breaths and relax to stablize blood pressure; also 
sit quietly for about 4-5 minutes before taking 
measurements.
or hot places; In a moving vehicle.
cannot be correct.
and black tea.
measurement.
the measurement.
electromagnetic radiation during the measurement
measurements  should be space 2-3 minutes)
the measurement results
Operation Instruction
Manual
Be prepared according to the correct method descriped 
earlier. 
Click “     ” button, the displayed by default will show the 
last user of the measured data after whole was out. 
Click: “     ” button, the display shows “0 mmHg” , 
followed by an air pump to inflate the cuff automatically, 
after the end of discouraged inflatable exhaust valve 
will automatically ladder, in this process, blood 
pressure and pulse rate can be measured.   
When measurement has been completed, the blood 
pressure and pulse data will be displayed on screen.
Take off the cuff, press “     ” button for 5 seconds, the 
monitor will be turn off. 
Note: When the monitor working, the “    ” symbol is 
always flashing. 
Emergency stop during measurement
During the measurement, if the user feels discomfort or 
other reasons need to interrupt the measurement, you can 
hold down the “     ” button for 5 seconds and then release 
it, then the monitor will go off and auto exhaust, and also 
can remove the cuff directly.   
Instruction of turn off automatically function
The monitor with auto turn off function, when the monitor 
will turn off automatically after keep 2 minutes still status 
at the symbol “    ” flashing.  
Instruction of battery status indicator
The monitor built-in rechargeable lithium battery, when 
the battery power is less than the preset value monitor, 
the battery symbol “    ” will be displayed as “red”. If the 
default static in the boot under the battery symbol is 
displayed as “red”, you need to immediately USB charging 
cable and output in the box for DC 5V 1A mobile phone 
charger is connected, and then charging cable's small 
plug into the machine's USB port for charging, when 
charging the battery symbo is displayed as “white” and 
“flicker” when the battery symbol to “Always” (about two 
hours), indicates that the battery is already full.        
Caution:  1.Continuous charging cannot more than a 
week, since overcharging may shorten 
battery life.
2.If the product willnot be used for a long time, 
battery life may be shortened and reduced 
performance, it may cause the battery to 
malfunction or cause a fire or explosion.
Please read  the care and maintenance
3.Please make the USB rubber cover well 
kept.
Using the same arm to measure your blood pressure at 
the same time every day
Please do not let the cuff is inflated long-term
1~2cm
9
中度⾼⾎压
重度⾼⾎压
2.Operation/transportation/ storage environment
Operation 
Environment
Operation temperature : +5~+40℃,
Humidity: 15%~80% RH,
Atmospheric pressure:80~105 kPa, 
avoid electromagnetic interference, vibration, 
dust, noise and large environment
Transportation 
and storage 
environment
 conditions, and may shake, bump environment
Temperature: -20~+55℃,
Humidity: 15%~80% RH,
No cold agglutination phenomenon, 
Atmospheric pressure: 80~105 kPa, 
avoid high temperatures, high humidity, 
direct sunlight, dusty, and causticity gas
Guide of operation under electromagnetic environment
When the monitor works with other devices may cause 
electromagnetic interference,  if interference phenomena 
occur when you use them, you should check the work 
environment to identify sources of interference and take 
the following measures to eliminate the interference:

Changing the direction of the interference device
Increase the interference between the device and the 
monitor  
8
Product specification
1.Technical specification
7
6
Troubleshooting
The following table is a common fault and possible error 
codes will be displayed in the course may encounter, as 
well as causes, exclude the method, if they are consistent 
with the situation in use can be handled with reference to 
the appropriate method.    
Common fault
Causation
Solution
Press“    ”button, 
the monitor 
cannot work
Battery voltage is 
low or exhausted
Charging power
The cuff has been 
wrapped incorrectly
Incorrect position
unstable mood
Wear too thick 
clothes
Influenced by 
environmental factors
Please read the relative 
chapter careful
Note: Blood pressure is dynamic, so the 
results of each measurement has some 
differences
Fault code
Causation
Solution
Er1
Over pressure, measurement object measured 
data exceeds the measurement range
Er2
Er6
When measuring the 
detected signal is too 
weak
The cuff is wrapped 
too loosely or not 
wrapped around the 
upper arm
Er3
Measurement overtime
Check the cuff for 
leaks, then re-test 
after have a rest
Er4
Er5
Er7
Er8
System error
Restart the monitor, 
then re-test
Excessive movement 
(with signal interference, 
etc.)
Please stay quietly 
when measured
Air leakage
Check the cuff is leak 
air or not
Pressure error
Restart and re-test
Blood pressure 
value is 
abnormally high 
or low
Caution: Please consult the distributor when  faced with the 
above conditions or malfunction can not be excluded 
in the process of using the monitor.
Measurement
values are
different with
each
measurement
5
( mmHg )
( mmHg )
120
130
140
160
180
80
85
90
100 110
Max hypertension
Lowest hypotension
2
3
5
6
-The most appropriate blood pressure
(Target value)
-Normal blood pressure
-Normal hypertension
-Mild hypertension
-Severe hypertension
-Moderate hypertension
1
1
2
4
5
6
3
Blood pressure status classification standard
Below is World Health Organization (WHO), the 
International Society of Hypertension blood pressure 
(ISH) developed classification standards.  
Referring to WHO / ISH blood pressure classification
 (1999 revised)
Hypotension has no clear definition, generally think that 
the systolic blood pressure less than 100mmHg 
(13.3kPa) for hypotension.
Note: Above symbol is for reference only, please 
be sure to accept the doctor's guidance.
4
Bluetooth status 
indicator
Product specification
Care and Maintenance
Product specification
Blood pressure
: ±3 mmHg(±0.4kPa)
Measurement  Blood pressure: 20~300 mmHg
(2.7~40 kPa)
MX-724BT
Product Name  Upper Arm Blood Pressure Monitor
Product Model 
Method of 
Oscillographic system
Display
Digital LED
Accuracy
Pulse: ±5% of display reading
Pulse: 40~199/min
Equipment 
Intermittent operation mode
Electrical 
Symbol Means BF Application
Symbol Means class II equipment
Device type
safety 
classification
operation 
mode
range
Measurement
No.
Pressure 
Resistive pressure sensor
Pressurization  Automatic pressurized inflation
Exhaust mode
Automatic pressure release, stepped deflation
Auto turn off
Measurement  From press “     ” button to start the measurement 
Battery
built-in 3.7V 800mAh rechargable Li battery
Method of 
With output DC5V/1A USB Cable for charging 
Charging time
About 2 hours
Weight
The main unit + cuff +Li battery total is approx: 
Dimension
Main unit: (Approx) 140.5 length x 56.3 width x 33 
Arm dimension  25~36cm 
Note: The above specifications are subject to change 
height (mm)
charging
period
to display the data needs 45 seconds
method
sensor
“    ”symbol is flash, the static time can be 2 
minutes
(Approx)263g
without notice.
To avoid damage to the unit by the accuracy and security 
measurements, please observe the following: 
The cuff balloon containing highly airtight, use caution, 
do not fold, pull, twist balloon. 
After each use, please keep the monitor and 
accessories tidy and properly placed, avoid strong 
force to the unit (such as dropping the unit on the floor).  
Don't put the monitor and accessories in high 
temperatures, high humidity, dust or direct sunlight 
environment.  
When cleaning the main unit, please wipe it with a soft, 
dry cloth, if necessary dampened with water or a 
neutral detergent, thoroughly wrung and then wipe, do 
not put it into the water to wash or use benzine, 
thinner, gasoline, etc. to wipe.    
Don't moisten the cuff or don't wash the cuff with water
Don't attempt to disassemble or repair the unit
 If the product is not used for above 6 month, please 
 charging the battery and place it very well. Each 6 
 month should check the power of the battery, if the 
 power is exhaust, charging it very well in time.   
Care and Maintenance
Portal 724, LLC
275 Hartz Way, Suite 105, Secaucus, NJ 07094 U.S.A
Note: Instructions for the folding tail A plane is the B-side of the head
Close the interference device 
-Consult the dealer or an experienced radio/TV technician 
 for help.
-Connect the equipment into an outlet on a circuit different 
 from that to which the receiver is connected.
-Increase the separatio between the equipment and 
 receiver.
-Reorient or telocate the receiving antenna.
If this equipment does cause harmful interference to radio 
or television, which can be determined by turning the 
equipment off and on, the user is encouraged to try to 
correct the interfrence by one or more of the following 
measures:
This device complies with Part 15 of the FCC Rules. 
Operation is subject to the following two conditions: 
(1) this device may not cause harmful interference  
(2) this device must accept any interference received, 
      including interference that may cause undesired 
      operation.
reasonable protection against harmful interference in a 
residential installation. This equipment generates, uses 
and can radiate radio frequency energy and, if not 
installed and used in accordance with the instructions, 
may cause harmful interference to radio communications. 
However, there is no guarantee that interference will not 
occur in a particular installation. 
device, pursuant to part 15 of the FCC Rules. These limits 
are designed to provide 
This equipment has been tested and found to comply with 
the limits for a Class B digital 
could void the user's authority to operate the equipment.
changes or modifications not expressly approved by the 
party responsible for compliance