Inspire Medical Systems Inc. 2500 Benutzerhandbuch

Seite von 39
26 English
Inspire Sleep Remote
200-161-101
Rev  A
3032PatEN_ch4.fm 10/30/15 8:55 am
4.625 x 6 inches (117 mm x 152 mm)     
Inspire Medical Confidential
Inspire 002
If you are having trouble positioning the sleep remote:
Note: 
These steps turn therapy off.
1. Make sure the audio tones are on. (For instructions on 
turning audio tones on and off, refer to page 23.)
2. Press the Therapy Off button (
).
3. Position the sleep remote directly over your generator 
and against your body. The buttons should face away 
from your body (Figure 12).
4. Adjust the sleep remote position until the Therapy On 
button (
) is directly over your generator. 
5. Wait until you hear a tone or 10 seconds pass.
6. Remove the sleep remote from your body and review the 
sleep remote lights to confirm that the change was 
communicated to your generator.
Notes:
• If you hear three rapid beeps, or the change you want is 
not indicated by the sleep remote lights, then try 
communicating with your generator again.
• The sleep remote communicates one change at a time. If 
you press an incorrect button, wait for 10 seconds until 
the sleep remote stops attempting to communicate with 
the generator, and then press the correct button. 
• Pressing the Therapy Off button (
) cancels 
communication in progress. The sleep remote then 
attempts to turn your generator off.