Adafruit DHT22 385 Benutzerhandbuch

Produktcode
385
Seite von 7
Aosong(Guangzhou) Electronics Co.,Ltd
---------------------------------------------------------------------------------------------------------------------
Tell: +86-020-36380552, +86-020-36042809
Fax: +86-020-36380562
Address: No.56, Renhe Road, Renhe Town, Baiyun District, Guangzhou, China
- 7 -
can recover after working in non-normal operating condition to calibrated status, but will accelerate sensors'
aging.
(2) Attentions to chemical materials
Vapor
from chemical materials may interfere AM2303's sensitive-elements and debase AM2303's sensitivity.
(3) Disposal when (1) & (2) happens
Step one: Keep the AM2303 sensor at condition of Temperature 50~60Celsius, humidity <10%RH for 2 hours;
Step two: After step one, keep the AM2303 sensor at condition of Temperature 20~30Celsius, humidity
>70%RH for 5 hours.
(4) Attention to temperature's affection
Relative humidity strongly depend on temperature, that is why we use temperature compensation technology
to ensure accurate measurement of RH. But it's still be much better to keep the sensor at same temperature when
sensing.
AM2303 should be mounted at the place as far as possible from parts that may cause change to temperature.
(5) Attentions to light
Long time exposure to strong light and ultraviolet may debase AM2303's performance.
(6) Attentions to connection wires
The connection wires' quality will effect communication's quality and distance, high quality shielding-wire is
recommended.
(7) Other attentions
* Welding temperature should be bellow 260Celsius.
* Avoid using the sensor under dew condition.
* Don't use this product in safety or emergency stop devices or any other occasion that failure of AM2303
may cause personal injury.