DS1000Z Programming Guide
:CURSor Commands
:CURSor commands are used to measure the X-axis value (such as time) and Y-axis value (such as voltage)
of the waveform displayed in the screen.
Command List:
:CURSor:MODE
Syntax
:CURSor:MODE <mode>
:CURSor:MODE?
:CURSor:MODE?
Description
Set or query the cursor measurement mode.
Parameter
Name
Type
Range
Default
<mode>
Discrete
{OFF|MANual|TRACk|AUTO|XY}
OFF
Explanation
OFF: disable the cursor measurement function.
MANual: enable the manual cursor measurement mode.
TRACk: enable the track cursor measurement mode.
AUTO: enable the auto cursor measurement mode.
XY: enable the XY cursor measurement mode. This mode is valid only when the
horizontal timebase mode is XY.
Return
Format
The query returns OFF, MAN, TRAC, AUTO or XY.
Example
:CURSor:MODE MANual /*enable the manual cursor mode*/
Related
Command