Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S Benutzerhandbuch
Produktcode
MSO1074Z-S
RIGOL
DS1000Z Programming Guide
2-59
:MEASure:STATistic:DISPlay
Syntax
:MEASure:STATistic:DISPlay <bool>
:MEASure:STATistic:DISPlay?
:MEASure:STATistic:DISPlay?
Description
Enable or disable the statistic function, or query the status of the statistic function.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{0|OFF}|{1|ON}}
0|OFF
Explanation
When the statistic function is enabled, the oscilloscope makes statistic and displays the
statistic results of at most 5 measurement items that are turned on last.
Return
Format
The query returns 0 or 1.
Example
:MEASure:STATistic:DISPlay ON
Related
Command
:MEASure:STATistic:MODE
Syntax
:MEASure:STATistic:MODE <mode>
:MEASure:STATistic:MODE?
:MEASure:STATistic:MODE?
Description
Set or query the statistic mode.
Parameter
Name
Type
Range
Default
<mode> Discrete
{DIFFerence|EXTRemum}
EXTRemum
Explanation
DIFFerence: select difference mode. The statistic results contain current value,
average value, standard deviation and counts.
EXTRemum: select extremum mode. The statistic results contain current value,
average value, minimum and maximum.
When the statistic function is enabled, the oscilloscope makes statistic and displays
the statistic results of at most 5 measurement items that are turned on last.
Return
Format
The query returns DIFF or EXTR.
Example
:MEASure:STATistic:MODE DIFFerence /*Select difference mode*/
:MEASure:STATistic:MODE? /*The query returns DIFF*/
Related
Command