Rigol DS2072A 2-channel oscilloscope, Digital Storage oscilloscope, DS2072A Benutzerhandbuch

Produktcode
DS2072A
Seite von 284
RIGOL 
  DS2000A User’s Guide 
5-30 
Trigger Condition: 
Press When to select the desired trigger condition. 
  >: trigger when the duration of the pattern is greater than the preset time. 
Press Time to set the duration of duration trigger and the range is from 2 ns to 
4 s. 
  <: trigger when the duration of the pattern is lower than the preset time. Press 
Time to set the duration of duration trigger and the range is from 2 ns to 4 s.   
  <>: trigger when the duration of the pattern is lower than the upper limit of the 
preset time and greater than the lower limit of the preset time. Press Upper 
Limit to set the upper limit of the duration of duration trigger and the range is 
from 12 ns to 4 s. Press Lower Limit to set the lower limit of the duration of 
duration trigger and the range is from 2 ns to 3.99 s. Note that the time lower 
limit must be lower than the time upper limit.   
 
Trigger Mode:   
Press Sweep to select the 
 (page 
) under this trigger type as auto, 
normal or single. The corresponding status light of the current trigger mode turns on. 
 
Trigger Setting: 
Press Setting to set the trigger parameters (trigger holdoff and noise rejection) 
under this trigger type.   
 
Trigger Level:   
Use TRIGGER 
  LEVEL to modify the level. For details, please refer to the 
description of 
 on page 
.