Rigol DS2072A 2-channel oscilloscope, Digital Storage oscilloscope, DS2072A Benutzerhandbuch

Produktcode
DS2072A
Seite von 284
RIGOL 
  DS2000A User’s Guide 
10-4 
Record Constant On   
 
DS2000A provides record “Open” (record constant on) mode for waveform record. 
When this mode is enabled, the oscilloscope records the waveform continuously 
according to the current memory depth. The length of the waveform recorded is 
limited by the memory depth and the waveform data recorded before will be 
overwritten by the waveform currently recorded. In this mode, the other oscilloscope 
operations (except 
 as well a
, 
 in the sample system setting and 
 in the horizontal system setting) are not affected.   
 
Press Utility  Record  Mode and use 
  to select “Open”. At this point, the 
waveform record key 
  at the front panel turns red and the oscilloscope records 
the waveform continuously according to the current memory depth. The oscilloscope 
stops recording until users press RUN/STOP and at this point, users can play back 
or analyze the waveform.   
 
In constant on mode, the oscilloscope can capture the accidental abnormal signals 
during the adjustment. Following is an application example of the record constant on 
mode. Use DS2000A (CH1) to observe a pulse which might contain runt pulses. 
 
1.  Connect the pulse under test to CH1 of the oscilloscope. Adjust the oscilloscope 
to make it trigger stably.