DS2000A User’s Guide
 
 
5-25  
 
Delay Trigger (Option) 
 
Trigger when the time difference (△T) between the specified edges of source A and 
source B meets the preset time limit, as shown in the figure below. Note that edge A 
and edge B must be neighbouring edges.   
T
Source A
Source B
Edge A=
Edge B=
 
 
Trigger Type: 
Press Type to select “Delay”. At this point, the trigger setting information as shown 
in the figure below is displayed at the upper right corner of the screen. 
 
 
Source A: 
Press SourceA to select CH1 or CH2 as the trigger source of signal source A. 
 
Edge A: 
Press EdgeA to select the trigger edge type of signal source A in delay trigger. It can 
be set to the rising edge or falling edge. 
 
Source B: 
Press SourceB to select CH1 or CH2 as the trigger source of signal source B. 
 
Edge B: 
Press EdgeB to select the trigger edge type of signal source B in delay trigger. It can 
be set to the rising edge or falling edge.