DS2000A User’s Guide
 
 
5-29  
 
Duration Trigger (Option) 
 
Identify a trigger condition by looking for the duration of a specified pattern. This 
pattern is a logical “AND” combination of the two channels. Each channel can have a 
value of high (H), low (L) or don’t care (X). The instrument triggers when the 
duration (△T) of this pattern meets the preset time, as shown in the figure below.   
T
T
CH1=H
CH2=L
CH1
CH2
 
 
Trigger Type: 
Press Type to select “Duration”. At this point, the trigger setting information as 
shown in the figure below is displayed at the upper right corner of the screen.   
 
 
Source Selection:   
Press Source to select CH1 or CH2 as the channel source of H, L or X. The current 
signal source is displayed at the upper right corner of the screen. 
 
Pattern Setting: 
Press Code to set the pattern of the current channel. At this point, the pattern 
setting area (as shown in the figure below) is displayed at the bottom of the screen. 
 
 
: set the pattern of the channel selected to “H”, namely the voltage level is 
higher than the threshold level of the channel.   
 
: set the pattern of the channel selected to “L”, namely the voltage level is 
lower than the threshold level of the channel.   
 
: set the pattern of the channel selected to “Don’t Care”, namely this channel 
is not used as a part of the pattern. When both the channels in the pattern are 
set to “Don’t Care”, the oscilloscope will not trigger.