  DS2000A User’s Guide 
5-38 
SPI Trigger 
 
Trigger on the data pattern on the specified edge. When using SPI trigger, you need 
to specify the SCL and SDA data sources. Below is the sequence chart of SPI bus data 
transmission.   
 
Serial Data Line Buffer
(SPIxRXB)
Shift Register
SPIxSR
SPI Buffer
SPIxBUF
Serial Data Line Buffer
(SPIxRXB)
Shift Register
SPIxSR
Serial Transmission Buffer
(SPIxTXB)
SPI Buffer
SPIxBUF
Serial Transmission Buffer
(SPIxTXB)
 
 
 
 
Highest Bit
Lowest Bit
Highest Bit
Lowest Bit
SDIx
SDOx
SDIx
SDOx
Processor 1[SPI  Host]
Processor 2[SPI Slave]
SCKx
Serial Clock
SCKx
 
 
Trigger Type: 
Press Type to select “SPI”. At this point, the trigger setting information as shown in 
the figure below is displayed at the upper right corner of the screen.   
 
 
Source Selection: 
Press SCL and SDA to specify the data sources of SCL and SDA respectively. They 
can be set to CH1 or CH2 and the current trigger source is displayed at the upper 
right corner of the screen.   
 
Data Line Setting: 
Set the instrument to trigger when the specified bit and length of data is transmitted 
in the SDA. 
  Press Data Bits to set the number of bits of the serial data character string. It