Waveform Playback 
 
Waveform playback can play back the waveforms currently recorded. Press Utility 
 Record  Mode and use 
  to select “Play back”. At this point, the information 
as shown in the figure below is displayed at the upper right corner of the screen, 
indicating the specific frame currently displayed on the screen. During the playback, 
this value would change continuously.   
 
 
Please refer to the following descriptions to set the playback parameters. 
 
1.  Play Mode 
Press this softkey to set the playback mode to cycle or single. 
 
: cycle playback. Play from the start frame to the end frame and then 
repeat until you stop it manually. 
 
: single playback. Play from the start frame to the end frame and 
then stop. 
 
2.  Interval 
Press this softkey and use 
  to set the time interval of playback. The range 
availbale is from 100 ns to 10 s and the default is 100 ns. 
 
3.  Start Frame 
Press this softkey and use 
  to set the start frame of playbak. The default is 1 
and the maximum is the maximum number of frames recorded. 
 
4.  Current Frame 
Press this softkey and use 
  or the navigation knob (large scale adjustment) 
to set the current frame. By default, the current frame is the same with the end 
frame. The setting range of current frame is related to the start frame and end 
frame settings. During the setting, the screen will display the corresponding 
waveform of the current frame synchronously, namely manual playback.   
If you perform the playback operation after setting this parameter, this menu will 
be adjusted to the Start Frame value automatically and will change 
continuously during the playback process.