All the specifications are guaranteed except the parameters marked with “Typical” 
and the oscilloscope needs to operate for more than 30 minutes under the specified 
operation temperature.   
 
 
Sample 
Sample Mode 
Real-time Sample 
Real Time 
Sample Rate 
2 GSa/s (single-channel) 
1 Gsa/s (dual-channel)   
Peak Detect 
500 ps (single-channel) 
1 ns (dual-channel) 
Averaging 
After both the channels finish N samples at the same time, N can 
be 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024, 2048, 4096 or 8192. 
High 
Resolution 
12 bits of resolution when ≥5 μs/div @ 1 GSa/s (or ≥10 μs/div @ 
500 MSa/s). 
Memory Depth 
single-channel: Auto, 14k pts, 140k pts, 1.4M pts, 14M pts and 
56M pts (option) are available 
dual-channel: Auto, 7k pts, 70k pts, 700k pts, 7M pts and 28M pts 
(option) are available 
 
 
Input 
Number of 
Channels 
Two channels 
Input Coupling  DC, AC or GND 
Input 
Impedance   
(1 MΩ±1%) || (16 pF±3 pF) or 50 Ω±1.5% 
Probe 
Attenuation 
Coefficient   
0.01X to 1000X, in 1-2-5 step 
Maximum 
Input Voltage 
Maximum Input Voltage of the Analog Channel 
CAT I 300 Vrms, CAT II 100 Vrms,