DS2000A User’s Guide
 
 
5-13  
 
available is from 2 ns to 4 s.   
  <>: trigger when the runt pulse width is greater than the lower limit and lower 
than the upper limit of pulse width. Press Upper Limit to set the maximum 
pulse width of runt trigger and the range is from 10 ns to 4 s; press Lower 
Limit
 to set the minimum pulse width of runt trigger and the range is from 2 ns 
to 3.99 s.   
Note: The lower limit of the pulse width must be lower than the upper limit. 
 
Vertical Window: 
Press Vertical to select the desired vertical window type. Note that under the “Runt 
trigger” menu, you can press the trigger level knob continuously to switch among 
different vertical window types. 
You can select the boundary of the trigger level to be set and then use Trigger   
LEVEL
 to adjust the trigger level. During the adjustment, two orange trigger level 
lines and trigger marks (
  and 
) appear on the screen and move up and down 
with the rotation of the knob. At the same time, the current trigger level values are 
displayed at the lower left corner of the screen. The trigger level lines and trigger 
marks disappear after you stop rotating the knob for about 2 s. 
 
 
 
The adjustment mode of the trigger level differs when different vertical window is 
selected.   
 
: only adjust the upper limit of the trigger level. During the adjustment, 
the “Up Level” changes accordingly and “Low Level” remains unchanged.   
 
: only adjust the lower limit of the trigger level. During the adjustment, 
the “Low Level” changes accordingly and the “Up Level” remains unchanged.   
 
: adjust the upper and lower limits of the trigger level at the same time. 
During the adjustment, the “Up Level” and “Low Level” change accordingly.   
 
Trigger Mode:   
Press Sweep to select the 
 (page 
) under this trigger type as auto, 
normal or single. The corresponding status light of the current trigger mode turns on. 
 
Trigger Setting:   
Press Setting to set the trigger parameters (trigger holdoff and noise rejection) 
under this trigger type.