Rigol DS2202A 2-channel oscilloscope, Digital Storage oscilloscope, DS2202A Benutzerhandbuch

Produktcode
DS2202A
Seite von 284
RIGOL 
  DS2000A User’s Guide 
5-16 
Nth Edge Trigger (Option)   
 
Trigger on the nth edge that appears after the specified idle time, as shown in the 
figure below.   
P1
P2
P3
M
P1/P2/P3<Idle Time<M
Edge Type=
Edge Num=2
 
 
Trigger Type: 
Press Type to select “Nth Edge”. At this point, the trigger setting information as 
shown in the figure below is displayed at the upper right corner of the screen.   
   
 
Source Selection: 
Press Source  to select CH1 or CH2 as the 
. The current trigger 
source is displayed at the upper right corner of the screen. 
Note: Select channel with signal input as trigger source to obtain stable trigger. 
 
Edge Type: 
Press Slope to select the kind of edge of the input signal on which the oscilloscope 
triggers. 
 
: trigger on the rising edge of the input signal when the voltage level meets 
the specified trigger level. 
 
: trigger on the falling edge of the input signal when the voltage level meets 
the specified trigger level. 
 
Idle Time: 
Press Idle to set the idle time before the edge counting in Nth egde trigger. The 
range available is from 16 ns to 4 s.