DS2000A User’s Guide
 
 
7-7  
 
5.  Data Packet Setting 
As mentioned before, in RS232, you need to set the start bit, data bits, check bit 
(optional) and stop bit of each frame of data. “Start Bit” is specified by the 
“Polarity Setting”. The setting methods of other parameters are as follows. 
  Press Data Bits to set the data width of each frame. It can be set to 5, 6, 7, 
8 or 9 and the default is 8. 
  Press Stop Bit to set the stop bit after each frame of data. It can be set to 
1 bit, 1.5 bits or 2 bits. 
  Press Even-OddCheck to set the even-odd check mode of the data 
transmission. It can be set to None, Odd Checkout or Even Checkout. 
  Press Packet to enable or disable the packet end. When packet end is 
enabled, several data blocks are combined according to the packet end.   
  Press PacketEnd to set the packet end during data transmission and it can 
be set to 00 (NULL), 0A (LF), 0D (CR), 20 (SP) or FF. 
 
6.  Display-related Setting 
Press Format to set the display format of the bus to Hex, Decimal, Binary or 
ASCII. 
Press Offset and use 
  to adjust the vertical display position of the bus. 
Press BusStatus to turn the bus display on or off. 
 
7.  Decoding Table 
The decoding table displays the decoded data, the corresponding line number, 
time and error information on TX and RX data lines in table format. It can be 
used to observe relatively longer decoded data to solve the problem that some 
data could not be seen clearly on the screen. Note that when the RX channel is 
turned off, the information of this data line will not be displayed in the decoding 
table.   
 
Press Event Table  Event Table to select “ON” (note that this operation is 
only available when BusStatus is set to “ON”) to enter the decoding table 
interface as shown in the figure below. The decoding table lists the decoded 
data in time order. If error occurs during the decoding, the corresponding error 
information is displayed. If a USB storage device is currently connected to the 
instrument, press Export to export the data table to the external USB storage 
device in CSV format.