Rigol DS2102A 2-channel oscilloscope, Digital Storage oscilloscope, DS2102A Benutzerhandbuch

Produktcode
DS2102A
Seite von 284
RIGOL 
  DS2000A User’s Guide 
10-10 
4.  Cancel   
During the analysis, users can press Cancel to stop the analysis and press Start 
again to restart the analysis.   
 
5.  Previous   
After the waveform analysis finishes, pressing Previous can locate the error 
frame previous to the current error frame. Pressing RUN/STOP can also 
perform the operation. 
 
6.  Next   
After the waveform analysis finishes, pressing Next can locate the error frame 
following the current error frame. Pressing SINGLE can also perform the 
operation. 
 
7.  Current Frame   
Press Current Frame to adjust the data frame currently displayed and the 
adjustable range is from 1 to the total number of frames in this waveform 
analysis. Using the navigation knob can also perform the operation. 
 
8.  Setup   
Press Setup to open the more detailed setting menu.   
  Screen Start: press this key to set the start point of waveform analysis and 
the range is from 5 to 685. The start point must be less than the “Screen 
End - 10” currently set.   
  Screen End: press this key to set the end point of waveform analysis and 
the range is from 15 to 695. The end point must be greater than the 
“Screen Start + 10” currently set.   
  Start Frame: press this key to set the start frame of waveform analysis and 
the default is the first frame.   
  End Frame: press this key to set the end frame of waveform analysis and 
the default is the last frame.   
  Threshold: press this key to set the threshold of waveform analysis and the 
range is from 1% to 99%. The threshold is used to judge whether the frame 
is an error frame. A frame is recognize as an error one if the (relative) 
difference value between the frame and the templet is equal to or larger 
than the threshold currently set.