Rigol DS2102A 2-channel oscilloscope, Digital Storage oscilloscope, DS2102A Benutzerhandbuch

Produktcode
DS2102A
Seite von 284
 
RIGOL 
DS2000A User’s Guide
 
 
10-13  
 
Analysis Based on Pass/Fail Mask 
 
Press Analyze and select “Pass/Fail”. Then, set the templet used in analysis based 
on Pass/Fail mask through the method below.   
 
Press MaskRange to open the following setting menus. 
 
1.  Mask Range 
Press this softkey to select “Screen Region” or “Cursor Region” for the analysis 
(the default is “Screen Region”). When “Cursor Region” is selected, two gray 
cursor lines appear on the screen. At this point, press Cursor A and Cursor B
then, use 
  to adjust the positions of the two cursor lines respectively to 
determine the analysis range. Or, press CursorAB and use 
  to adjust the 
positions of cursor A and curcor B at the same time. Note that you can press 
down 
  continuously to switch the current cursor. 
 
2.  X Mask   
Press this softkey and use 
  to adjust the horizontal threshold and the range 
is from 0.02 div to 4.00  div.  During the adjustment, two curves would be 
displayed to show the outline of the current mask as shown in the figure below.   
 
 
 
3.  Y Mask 
Press this softkey and use 
  to adjust the vertical threshold and the range is 
from 0.04 div to 5.12 div. During the adjustment, two curves would be displayed 
to show the outline of the current mask. 
 
Outline of the Current Mask