After selecting FM, you can set the FM frequency and frequency deviation of 
the modulating waveform.   
 
FM Frequency: press FM_Freq to set the frequency of the modulating 
waveform. The range is from 1 Hz to 50 kHz. For the setting method, please 
refer to the introduction in 
. 
 
Frequency Deviation: press Deviation to set the deviation of the 
frequencyof the modulating waveform relative to the carrier frequency. The 
range is from 0 Hz to the current carrier frequency.   
 
2.  Select modulating waveform 
Press Shape to select sine, square, ramp waveform or noise as the modulating 
waveform.   
Modulating Waveform 
Carrier Waveform 
Modulated Waveform