Rigol DS2102A 2-channel oscilloscope, Digital Storage oscilloscope, DS2102A Benutzerhandbuch

Produktcode
DS2102A
Seite von 284
RIGOL 
  DS2000A User’s Guide 
4-8 
Memory Depth 
 
Memory depth refers to the number of waveform points that the oscilloscope can 
store in a single trigger sample and it reflects the storage ability of the sample 
memory. DS2000A provides up to 56 Mpts memory depth (option) and 14 Mpts 
standard memory depth.   
 
T
Trigger Point
Pre-sample
Delayed Sample
Memory Depth
 
 
The relation of memory depth, sample rate and waveform length fulfills the equation 
below:   
 
Memory depth = sample rate (Sa/s) × waveform length (s/div × div) 
 
Therefore, under the same timebase scale, higher memory depth can ensure higher 
sample rate.   
 
Press Acquire  Mem Depth, use 
  to switch to the desired memory depth (the 
default is auto) and then press down the knob to select the option. You can also 
press Mem Depth continuously to switch the memory depth. 
 
When a single channel is enabled, the memory depths available include Auto, 
14kPoints, 140kPoints, 1.4MPoints, 14MPoints and 56MPoints (option). In “Auto” 
mode, the oscilloscope selects the memory depth automatically according to the 
current sample rate.   
 
When CH1 and CH2 are enabled, the memory depths available include Auto, 7kPoints,