Trigger Mode 
 
Trigger mode affects the way in which the oscilloscope searches for the trigger. The 
following is the schematic diagram of the acquisition memory. As shown in the figure 
below, the position of the trigger event is determined by the reference time point and 
the delay setting. Note that the acquisition memory of the oscilloscope is a cyclic 
buffer and the new data would overwrite the old data until the acquisition finishes. 
 
 
 
 
 
 
 
 
 
Pre-trigger/Delayed trigger: 
Acquire data before/after the trigger event. The trigger position is usually at the 
horizontal center of the screen. In full-screen display, seven-grid pre-trigger and 
delayed trigger information are displayed respectively. You can adjust the horizontal 
position of the waveform through HORIZONTAL 
  POSITION to view more 
pre-trigger information and delayed trigger information, through which the signal 
information before/after the trigger (such as capture the glitch generated by the 
circuit and analyze the pre-trigger data to find out the reasons for glitch) can be 
obtained.   
 
Press MODE in the trigger control area (TRIGGER) at the front panel or press 
MENU  Sweep to select the desired trigger mode. The corresponding status light 
of the mode currently selected turns on.   
 
Auto: 
No matter whether the trigger condition is met, there is always waveform display. A 
horizontal line is displayed when no signal is input.     
In this mode, the oscilloscope operates by first filling the pre-trigger buffer. It starts 
searching for a trigger after the pre-trigger buffer is filled and continues to flow data 
through this buffer while it searches for the trigger. While searching for the trigger, 
Pre-trigger Buffer 
Post-trigger Buffer 
Trigger Event 
Acquisition memory