5-6 
Trigger Holdoff   
 
Trigger holdoff can be used to stably trigger the complex waveforms (such as pulse 
series). Holdoff time is the amount of time that the oscilloscope waits before 
re-arming the trigger circuitry. The oscilloscope will not trigger until the holdoff time 
expires.   
 
 
 
 
 
Press MENU  Setting  Holdoff in the trigger control area (TRIGGER) at the 
front panel and use 
  to modify the holdoff time (the default is 100 ns) until the 
waveform triggers stably. The adjustable range of holdoff time is from 100 ns to 10 s. 
Note that trigger holdoff is not available for Nth edge trigger, video trigger, RS232 
trigger, I2C trigger, SPI trigger and USB trigger.   
 
 
 
 
 
 
 
 
 
 
 
 
Holdoff Time          
 
Trigger Position