Windows Trigger (Option) 
 
Windows trigger provides a high trigger level and a low trigger level. The instrument 
triggers when the input signal passes through the high trigger level or the low trigger 
level.   
 
Trigger Type: 
Press Type to select “Windows”. At this point, the trigger setting information as 
shown in the figure below is displayed at the upper right corner of the screen.   
 
 
Source Selection: 
Press Source to select CH1 or CH2 as the 
. The current trigger 
source is displayed at the upper right corner of the screen. 
Note: Select channel with signal input as trigger source to obtain stable trigger. 
 
Windows Type: 
Press WndType to select the kind of edge of the input signal on which the 
oscilloscope triggers. 
 
: trigger on the rising edge of the input signal when the voltage level is 
greater than the preset high trigger level.   
 
: trigger on the falling edge of the input signal when the voltage level is 
lower than the preset low trigger level.   
 
: trigger on any edge of the input signal when the voltage level meets the 
preset trigger level. 
 
Trigger Position: 
After selecting the windows type, press Position to further specify the time point of 
trigger by selecting the trigger position.   
  Enter: trigger when the trigger signal enters the specified trigger level range.   
  Exit: trigger when the input signal exits the specified trigger level range.   
  Time: used to specify the hold time of the input signal after entering the 
specified trigger level range. The instrument triggers when the accumulated 
hold time equals the windows time.