  DS2000A User’s Guide 
12-6 
To Output Noise 
 
Press Src1 Setup to open the waveform setting menu. Press Wave and select 
“Noise”. At this point, you can output noise signal with specified amplitude.   
 
1.  Source 
Press Source to turn on or off the signal output. 
 
2.  Set the amplitude 
Press Amplitude to set the amplitude of the signal. When the impedance is set 
to HighZ, the range is from 20 mVpp to 5 Vpp; when the impedance is set to 50 
Ω, the range is from 10 mVpp to 2.5 Vpp. For the setting method, please refer to 
the introduction i
.   
 
3.  Set the offset 
Press Offset to set the DC offset of the noise signal. When the impedance is 
HighZ, the range is from (-2.5 V+the current amplitude/2) to (2.5 V-the current 
amplitude/2); when the impedance is set to 50 Ω, the range is from (-1.25 V+the 
current amplitude/2) to (1.25 V-the current amplitude/2).   
 
Method 1: 
Rotate 
  or the navigation knob to set the desired DC offset voltage. 
 
Method 2: 
Press down 
  and the numeric keyboard is displayed. Use the knob to input 
the desired offset value and unit.   
Note: The offset value contains three effective digits. For example, when the 
amplitude is set to 10 mVpp (the impedance is 50 Ω), the offset range calculated 
is from -1.245 V to 1.245 V, while the actual offset range is -1.24 V to 1.24 V.