 Segmented memory acquisition example: 
Segment
n
Segment
n+1
Segment
n+2
Segment
n+3
Segment
n+4
Segment
n+5
 
As shown below, the memory is divided into segments to increase 
the number of events that can be effectively captured with the same 
acquisition memory. Also notice that the scope doesn’t need to 
rearm the trigger between each segment, this makes the segmented 
memory function especially useful for high speed signals. The time 
between each segment is also recorded so that accurate signal 
timing can also be measured. 
The segmented memory function also supports automatic 
measurements for each segment or statistics for all the captured 
segments. 
The Advanced Segment Memory Utility is applicable for both 
analog and digital channels.