Gw Instek GDS-2302A 2-channel oscilloscope, Digital Storage oscilloscope, GDS-2302A Datenbogen

Produktcode
GDS-2302A
Seite von 261
 
 
GDS-2000A Series User Manual 
148
 
Coupling 
(Edge, Delay) 
DC 
DC coupling.  
AC 
AC coupling. Blocks DC components 
from the trigger circuits. 
HF reject 
High frequency filter above 100kHz  
 
LF reject 
Low frequency filter below 5kHz 
 
Reject noise 
DC coupling with low sensitivity to 
reject noise. 
Slope 
(Edge, Delay, Rise 
& Fall) 
 
Trigger on a rising edge. 
 
Trigger on a falling edge. 
 
Either. (either rising or falling edge) 
(Edge, Delay, Rise & Fall trigger type 
only) 
Trigger Level 
(Edge, Delay) 
Level 
Adjusts the trigger 
manually using the 
Trigger LEVEL knob. 
LEVEL
 
 
Set to TTL 
1.4V 
Sets the trigger level to 1.4V, suitable 
for triggering on TTL signals. 
 
Set to ECL -
1.3V 
Sets the trigger to -1.3V. This is 
suitable for ECL circuits. 
 
Set to 50% 
Sets the trigger level to 
50% of the waveform 
amplitude.  
50 %
 
Holdoff 
Holdoff 
Sets the holdoff time.  
Set to 
Minimum 
Set the holdoff time to the minimum.  
Delay  
(Delay) 
Time 
Sets the delay time (10ns ~ 10s) 
between the trigger event and the real 
trigger timing.